Validation of a fully autonomous phosphate analyser based on a microfluidic lab-on-a-chip

C. Slater*, J. Cleary, K. T. Lau, D. Snakenborg, B. Corcoran, J. P. Kutter, D. Diamond

*Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

29 Citations (Scopus)

Abstract

This work describes the design of a phosphate analyser that utilises a microfluidic lab-on-a-chip. The analyser contains all the required chemical storage, pumping and electronic components to carry out a complete phosphate assay. The system is self-calibrating and self-cleaning, thus capable of long-term operation. This was proven by a bench top calibration of the analyser using standard solutions and also by comparing the analyser's performance to a commercially available phosphate monitor installed at a waste water treatment plant. The output of the microfluidic lab-on-a-chip analyser was shown to have sensitivity and linear range equivalent to the commercially available monitor and also the ability to operate over an extended period of time.